Electro Technical Calibration

The Electro-Technical Calibration Division at the Central Electronics Centre (CEC), IIT Madras offers precision calibration services for a broad range of electrical and electronic instruments. All calibrations are traceable to national and international standards, and are performed using high-accuracy reference instruments like multiproduct calibrators, current coils, digital multimeters (up to 8.5-digit resolution), Decade Resistance box and Precision current shunt. Focusing on reliability and regulatory compliance, the division serves academic, industrial, and research stakeholders by maintaining high standards of uncertainty, repeatability, and traceability.

CEC’s Electro-Technical Calibration delivers high-accuracy, traceable measurements across voltage, current, power, resistance, frequency, and time domains.

CEC’s electro-technical calibration capabilities span AC and DC current (1 µA to 1000 A), AC and DC voltage (100 µV to 1000 V), power measurements (AC power from 1 W to 12 kW and DC power from 10 mW to 20 kW), and power factor calibration (0.2 PF to 1 PF). Resistance calibrations are supported across a wide range from 1 mΩ to 1 TΩ. The division also calibrates frequency devices (from 1 Hz to 500 MHz), timer(from 1s to 24 hrs) and provides comprehensive oscilloscope calibration of bandwidth up to 600 MHz, and time base resolution down to 2 ns. In Measure Mode, CEC simulates temperature sensor outputs and verifies the response of thermocouples and RTDs. CMC values range from ±0.016 °C to ±0.09 °C, depending on the sensor type and range. In Source Mode, simulated thermoelectric signals are applied to instruments such as indicators and DAQs to validate performance, with accuracies as tight as ±0.06 °C.

International Standards Compliance

Calibrations are traceable to EURAMET Cg-15 and Cg-7, ensuring global reliability and regulatory acceptance.

Extensive Calibration Ranges

Supports ultra-low to high-level calibrations across electrical quantities — including voltage, current, resistance, and frequency.

High-End Reference Equipment

Equipped with multiproduct calibrators, high-voltage sources, and 8.5-digit multimeters for top-tier accuracy and precision

Broad Device Compatibility

Supports calibration for DMMs, oscilloscopes, power analyzers, frequency counters, and more — all under one roof.

Electro Technical Calibration: Measure Mode

Sno
Parameter
Measurand
Range
CMC (±)
Calibration Standard Equipment
Device under Calibration (DUC)
1
AC Current
at 50 Hz
20 A to 900 A
0.31% to 0.48%
8.5-digit multimeter, HV probe, Current shunt, Timer
AC/DC voltage sources, AC/DC Current sources, AC/DC power supplies, High voltage AC/DC sources, Programmable AC/DC power supplies, HV AC testers, HV DC Generators, HV DC current sources, Decade Resistance Box, Timer, Power Analyzer, Function/Arbitrary Waveform Generator.
1
AC Current
50 Hz to 1 kHz
1 uA to 20 A
2.42% to 0.12%
8.5-digit multimeter, HV probe, Current shunt, Timer
AC/DC voltage sources, AC/DC Current sources, AC/DC power supplies, High voltage AC/DC sources, Programmable AC/DC power supplies, HV AC testers, HV DC Generators, HV DC current sources, Decade Resistance Box, Timer, Power Analyzer, Function/Arbitrary Waveform Generator.
1
AC Current
50 Hz to 10 kHz
1 mA to 1 A
0.059% to 0.12%
8.5-digit multimeter, HV probe, Current shunt, Timer
AC/DC voltage sources, AC/DC Current sources, AC/DC power supplies, High voltage AC/DC sources, Programmable AC/DC power supplies, HV AC testers, HV DC Generators, HV DC current sources, Decade Resistance Box, Timer, Power Analyzer, Function/Arbitrary Waveform Generator.
2
AC Voltage
10 Hz to 1 kHz
1 mV to 1000 V
0.914% to 0.007%
8.5-digit multimeter, HV probe, Current shunt, Timer
AC/DC voltage sources, AC/DC Current sources, AC/DC power supplies, High voltage AC/DC sources, Programmable AC/DC power supplies, HV AC testers, HV DC Generators, HV DC current sources, Decade Resistance Box, Timer, Power Analyzer, Function/Arbitrary Waveform Generator.
2
AC Voltage
1 kHz to 100 kHz
1 mV to 100 V
1.32% to 0.008%
8.5-digit multimeter, HV probe, Current shunt, Timer
AC/DC voltage sources, AC/DC Current sources, AC/DC power supplies, High voltage AC/DC sources, Programmable AC/DC power supplies, HV AC testers, HV DC Generators, HV DC current sources, Decade Resistance Box, Timer, Power Analyzer, Function/Arbitrary Waveform Generator.
2
AC Voltage
1 kHz to 20 kHz
100 V to 1000 V
0.008% to 0.007%
8.5-digit multimeter, HV probe, Current shunt, Timer
AC/DC voltage sources, AC/DC Current sources, AC/DC power supplies, High voltage AC/DC sources, Programmable AC/DC power supplies, HV AC testers, HV DC Generators, HV DC current sources, Decade Resistance Box, Timer, Power Analyzer, Function/Arbitrary Waveform Generator.
2
AC Voltage
100 kHz to 1 MHz
1 mV to 10 V
1.38% to 0.14%
8.5-digit multimeter, HV probe, Current shunt, Timer
AC/DC voltage sources, AC/DC Current sources, AC/DC power supplies, High voltage AC/DC sources, Programmable AC/DC power supplies, HV AC testers, HV DC Generators, HV DC current sources, Decade Resistance Box, Timer, Power Analyzer, Function/Arbitrary Waveform Generator.
3
AC High Voltage
at 50 Hz
1 kV to 28 kV
1.2% to 5.77%
8.5-digit multimeter, HV probe, Current shunt, Timer
AC/DC voltage sources, AC/DC Current sources, AC/DC power supplies, High voltage AC/DC sources, Programmable AC/DC power supplies, HV AC testers, HV DC Generators, HV DC current sources, Decade Resistance Box, Timer, Power Analyzer, Function/Arbitrary Waveform Generator.
4
DC Current
-
1 uA to 800 A
0.62% to 0.21%
8.5-digit multimeter, HV probe, Current shunt, Timer
AC/DC voltage sources, AC/DC Current sources, AC/DC power supplies, High voltage AC/DC sources, Programmable AC/DC power supplies, HV AC testers, HV DC Generators, HV DC current sources, Decade Resistance Box, Timer, Power Analyzer, Function/Arbitrary Waveform Generator.
5
DC Voltage
-
10 uV to 1000 V
1.42% to 0.001%
8.5-digit multimeter, HV probe, Current shunt, Timer
AC/DC voltage sources, AC/DC Current sources, AC/DC power supplies, High voltage AC/DC sources, Programmable AC/DC power supplies, HV AC testers, HV DC Generators, HV DC current sources, Decade Resistance Box, Timer, Power Analyzer, Function/Arbitrary Waveform Generator.
6
DC High Voltage
-
1 kV to 40 kV
1.2% to 2.31%
8.5-digit multimeter, HV probe, Current shunt, Timer
AC/DC voltage sources, AC/DC Current sources, AC/DC power supplies, High voltage AC/DC sources, Programmable AC/DC power supplies, HV AC testers, HV DC Generators, HV DC current sources, Decade Resistance Box, Timer, Power Analyzer, Function/Arbitrary Waveform Generator.
7
Resistance
2 wire
100 uΩ to 100 MΩ
0.13% to 0.007%
8.5-digit multimeter, HV probe, Current shunt, Timer
AC/DC voltage sources, AC/DC Current sources, AC/DC power supplies, High voltage AC/DC sources, Programmable AC/DC power supplies, HV AC testers, HV DC Generators, HV DC current sources, Decade Resistance Box, Timer, Power Analyzer, Function/Arbitrary Waveform Generator.
7
Resistance
2 wire
10 MΩ to 10 GΩ
0.004% to 0.29%
8.5-digit multimeter, HV probe, Current shunt, Timer
AC/DC voltage sources, AC/DC Current sources, AC/DC power supplies, High voltage AC/DC sources, Programmable AC/DC power supplies, HV AC testers, HV DC Generators, HV DC current sources, Decade Resistance Box, Timer, Power Analyzer, Function/Arbitrary Waveform Generator.
7
Resistance
4 wire
1 Ω to 10 MΩ
0.006% to 0.004%
8.5-digit multimeter, HV probe, Current shunt, Timer
AC/DC voltage sources, AC/DC Current sources, AC/DC power supplies, High voltage AC/DC sources, Programmable AC/DC power supplies, HV AC testers, HV DC Generators, HV DC current sources, Decade Resistance Box, Timer, Power Analyzer, Function/Arbitrary Waveform Generator.
8
Time
1 s to 86400 s
0.062 s to 14.21 s
8.5-digit multimeter, HV probe, Current shunt, Timer
AC/DC voltage sources, AC/DC Current sources, AC/DC power supplies, High voltage AC/DC sources, Programmable AC/DC power supplies, HV AC testers, HV DC Generators, HV DC current sources, Decade Resistance Box, Timer, Power Analyzer, Function/Arbitrary Waveform Generator.
Sno
Parameter
Measurand
Range
CMC (±)
Calibration Standard Equipment
Device under Calibration (DUC)
1
Temperature Simulation
B Type Thermocouple
600°C to 1800°C
0.08°C
Temperature Simulation /
8.5 digit Multimeter
Process Calibrator / Multiproduct Calibrator/ Multifunction Calibrator
2
Temperature Simulation
E Type Thermocouple
(-) 200°C to 1000°C
0.06°C
Temperature Simulation /
8.5 digit Multimeter
Process Calibrator / Multiproduct Calibrator/ Multifunction Calibrator
3
Temperature Simulation
J Type Thermocouple
(-) 200°C to 1200°C
0.06°C
Temperature Simulation /
8.5 digit Multimeter
Process Calibrator / Multiproduct Calibrator/ Multifunction Calibrator
4
Temperature Simulation
K Type Thermocouple
(-) 200°C to 1350°C
0.06°C
Temperature Simulation /
8.5 digit Multimeter
Process Calibrator / Multiproduct Calibrator/ Multifunction Calibrator
5
Temperature Simulation
N Type Thermocouple
(-) 200°C to 1300°C
0.06°C
Temperature Simulation /
8.5 digit Multimeter
Process Calibrator / Multiproduct Calibrator/ Multifunction Calibrator
6
Temperature Simulation
PT100 (385)
(-) 200°C to 800°C
0.016°C
Temperature Simulation /
8.5 digit Multimeter
Process Calibrator / Multiproduct Calibrator/ Multifunction Calibrator
7
Temperature Simulation
R Type Thermocouple
100°C to 1750°C
0.09°C
Temperature Simulation /
8.5 digit Multimeter
Process Calibrator / Multiproduct Calibrator/ Multifunction Calibrator
8
Temperature Simulation
S Type Thermocouple
100°C to 1750°C
0.08°C
Temperature Simulation /
8.5 digit Multimeter
Process Calibrator / Multiproduct Calibrator/ Multifunction Calibrator
9
Temperature Simulation
T Type Thermocouple
(-) 200°C to 400°C
0.08°C
Temperature Simulation /
8.5 digit Multimeter
Process Calibrator / Multiproduct Calibrator/ Multifunction Calibrator

Electro Technical Calibration: Source Mode

Sno
Parameter
Measurand
Range
CMC (±)
Calibration Standard Equipment
Device under Calibration (DUC)
1
AC Current
at 50 Hz
20 A to 1000 A
0.31% to 0.48%
Multiproduct calibrator / Current coil / Standard DRB/ Decade Megaohms Resistance Box
AC/DC Clamp Meter, DMM(upto 6.5 digit), AC/DC Ammeter, AC/DC Voltmeter, Power Meter, Power Analyzer, Precision Shunt, Oscilloscope, Frequency Counter
1
AC Current
50 Hz to 1 kHz
1 uA to 20 A
0.61% to 0.34%
Multiproduct calibrator / Current coil / Standard DRB/ Decade Megaohms Resistance Box
AC/DC Clamp Meter, DMM(upto 6.5 digit), AC/DC Ammeter, AC/DC Voltmeter, Power Meter, Power Analyzer, Precision Shunt, Oscilloscope, Frequency Counter
2
AC Voltage
50 Hz to 1 kHz
100 V to 1000 V
0.039% to 0.046%
Multiproduct calibrator / Current coil / Standard DRB/ Decade Megaohms Resistance Box
AC/DC Clamp Meter, DMM(upto 6.5 digit), AC/DC Ammeter, AC/DC Voltmeter, Power Meter, Power Analyzer, Precision Shunt, Oscilloscope, Frequency Counter
2
AC Voltage
50 Hz to 10 kHz
1 mV to 100 V
1.34% to 0.046%
Multiproduct calibrator / Current coil / Standard DRB/ Decade Megaohms Resistance Box
AC/DC Clamp Meter, DMM(upto 6.5 digit), AC/DC Ammeter, AC/DC Voltmeter, Power Meter, Power Analyzer, Precision Shunt, Oscilloscope, Frequency Counter
2
AC Voltage
10 kHz to 100 kHz
1 mV to 100 V
2.4% to 0.29%
Multiproduct calibrator / Current coil / Standard DRB/ Decade Megaohms Resistance Box
AC/DC Clamp Meter, DMM(upto 6.5 digit), AC/DC Ammeter, AC/DC Voltmeter, Power Meter, Power Analyzer, Precision Shunt, Oscilloscope, Frequency Counter
3
AC Power
at 50 Hz (10 V to 600 V/0.1 A to 20 A / UPF)
1 W to 12 kW
0.15% to 0.19%
Multiproduct calibrator / Current coil / Standard DRB/ Decade Megaohms Resistance Box
AC/DC Clamp Meter, DMM(upto 6.5 digit), AC/DC Ammeter, AC/DC Voltmeter, Power Meter, Power Analyzer, Precision Shunt, Oscilloscope, Frequency Counter
4
Power Factor
at 50 Hz (230V / 5A / Lead & Lag)
0.2 PF to 1 PF
0.0148 PF to 0.002 PF
Multiproduct calibrator / Current coil / Standard DRB/ Decade Megaohms Resistance Box
AC/DC Clamp Meter, DMM(upto 6.5 digit), AC/DC Ammeter, AC/DC Voltmeter, Power Meter, Power Analyzer, Precision Shunt, Oscilloscope, Frequency Counter
5
DC Current
-
1 uA to 1000 A
0.61% to 0.4%
Multiproduct calibrator / Current coil / Standard DRB/ Decade Megaohms Resistance Box
AC/DC Clamp Meter, DMM(upto 6.5 digit), AC/DC Ammeter, AC/DC Voltmeter, Power Meter, Power Analyzer, Precision Shunt, Oscilloscope, Frequency Counter
6
DC Voltage
-
100 uV to 1000 V
1.4% to 0.003%
Multiproduct calibrator / Current coil / Standard DRB/ Decade Megaohms Resistance Box
AC/DC Clamp Meter, DMM(upto 6.5 digit), AC/DC Ammeter, AC/DC Voltmeter, Power Meter, Power Analyzer, Precision Shunt, Oscilloscope, Frequency Counter
7
DC Power
(100 mV to 1000 V/100 mA to 20 A)
10 mW to 20 kW
0.03% to 0.08%
Multiproduct calibrator / Current coil / Standard DRB/ Decade Megaohms Resistance Box
AC/DC Clamp Meter, DMM(upto 6.5 digit), AC/DC Ammeter, AC/DC Voltmeter, Power Meter, Power Analyzer, Precision Shunt, Oscilloscope, Frequency Counter
8
Resistance
2 wire
100 kΩ to 1 GΩ
0.5% to 1.8%
Multiproduct calibrator / Current coil / Standard DRB/ Decade Megaohms Resistance Box
AC/DC Clamp Meter, DMM(upto 6.5 digit), AC/DC Ammeter, AC/DC Voltmeter, Power Meter, Power Analyzer, Precision Shunt, Oscilloscope, Frequency Counter
8
Resistance
2 wire at 1000 V
10 MΩ to 1 TΩ
1.29% to 2.59%
Multiproduct calibrator / Current coil / Standard DRB/ Decade Megaohms Resistance Box
AC/DC Clamp Meter, DMM(upto 6.5 digit), AC/DC Ammeter, AC/DC Voltmeter, Power Meter, Power Analyzer, Precision Shunt, Oscilloscope, Frequency Counter
8
Resistance
4 wire
1 mΩ to 100 kΩ
9.3% to 0.0037%
Multiproduct calibrator / Current coil / Standard DRB/ Decade Megaohms Resistance Box
AC/DC Clamp Meter, DMM(upto 6.5 digit), AC/DC Ammeter, AC/DC Voltmeter, Power Meter, Power Analyzer, Precision Shunt, Oscilloscope, Frequency Counter
9
Oscilloscope
AC Voltage at 1 kHz
1 mV to 130 V
4.89% to 1.03%
Multiproduct calibrator / Current coil / Standard DRB/ Decade Megaohms Resistance Box
AC/DC Clamp Meter, DMM(upto 6.5 digit), AC/DC Ammeter, AC/DC Voltmeter, Power Meter, Power Analyzer, Precision Shunt, Oscilloscope, Frequency Counter
9
Oscilloscope
Bandwidth at 50 kHz to 600 MHz
5 mVpp to 5.5 Vpp
5.78%
Multiproduct calibrator / Current coil / Standard DRB/ Decade Megaohms Resistance Box
AC/DC Clamp Meter, DMM(upto 6.5 digit), AC/DC Ammeter, AC/DC Voltmeter, Power Meter, Power Analyzer, Precision Shunt, Oscilloscope, Frequency Counter
9
Oscilloscope
DC Voltage
1 mV to 130 V
3.72% to 1.01%
Multiproduct calibrator / Current coil / Standard DRB/ Decade Megaohms Resistance Box
AC/DC Clamp Meter, DMM(upto 6.5 digit), AC/DC Ammeter, AC/DC Voltmeter, Power Meter, Power Analyzer, Precision Shunt, Oscilloscope, Frequency Counter
9
Oscilloscope
Time base
2 ns to 5 s
0.16%
Multiproduct calibrator / Current coil / Standard DRB/ Decade Megaohms Resistance Box
AC/DC Clamp Meter, DMM(upto 6.5 digit), AC/DC Ammeter, AC/DC Voltmeter, Power Meter, Power Analyzer, Precision Shunt, Oscilloscope, Frequency Counter
10
Frequency
1 Hz to 500 MHz
0.006% to 0.0003%
Multiproduct calibrator / Current coil / Standard DRB/ Decade Megaohms Resistance Box
AC/DC Clamp Meter, DMM(upto 6.5 digit), AC/DC Ammeter, AC/DC Voltmeter, Power Meter, Power Analyzer, Precision Shunt, Oscilloscope, Frequency Counter
Sno
Parameter
Measurand
Range
CMC (±)
Calibration Standard Equipment
Device under Calibration (DUC)
1
Temperature Simulation
B Type Thermocouple
600°C to 1800°C
0.22°C
Multiproduct Calibrator
Temperature measurement with Indicators/ Controllers/ DAQ
2
Temperature Simulation
E Type Thermocouple
(-) 200°C to 1000°C
0.06°C
Multiproduct Calibrator
Temperature measurement with Indicators/ Controllers/ DAQ
3
Temperature Simulation
J Type Thermocouple
(-) 200°C to 1200°C
0.06°C
Multiproduct Calibrator
Temperature measurement with Indicators/ Controllers/ DAQ
4
Temperature Simulation
K Type Thermocouple
(-) 200°C to 1350°C
0.06°C
Multiproduct Calibrator
Temperature measurement with Indicators/ Controllers/ DAQ
5
Temperature Simulation
N Type Thermocouple
(-) 200°C to 1300°C
0.06°C
Multiproduct Calibrator
Temperature measurement with Indicators/ Controllers/ DAQ
6
Temperature Simulation
RTD PT100 (385)
(-) 200°C to 800°C
0.06°C
Multiproduct Calibrator
Temperature measurement with Indicators/ Controllers/ DAQ
7
Temperature Simulation
R Type Thermocouple
100°C to 1750°C
0.09°C
Multiproduct Calibrator
Temperature measurement with Indicators/ Controllers/ DAQ
8
Temperature Simulation
S Type Thermocouple
100°C to 1750°C
0.08°C
Multiproduct Calibrator
Temperature measurement with Indicators/ Controllers/ DAQ
9
Temperature Simulation
T Type Thermocouple
(-) 200°C to 400°C
0.08°C
Multiproduct Calibrator
Temperature measurement with Indicators/ Controllers/ DAQ